Figure 1: The Navigate Window showing the alignment mark in the sample base
NOTE: Ensure that the head is raised high enough not to hit the EC cell when moving the XY stage.
|
|
|
Figure 2: Click 2.2 Teach Blind Engage Position
NOTE: When the using the EC cell, only ± 2470 μm (7000 μm circular, ± 5000 μm for the Electrochemistry Cell) of XY stage motion is possible. If you move more than that when the microscope is focused on the EC sample, your microscope may hit the inside of the EC cell.
|
|
CAUTION: When moving the XY stage when the AFM microscope is focused on the EC sample, ensure that these coordinates are never greater than ± 2470 μm.
|
|
|
Figure 3: Click 3.1 Move to Add Fluid Position
|
|
NOTE: The Dimension General Purpose Electrochemistry Cell has a capacity of 2.5 ml of fluid. Do not overfill the EC cell or damage to the heater in the EC chuck may result. Ensure that the fluid level is sufficiently high so that it touches the bottom window of the probe holder.
|
|
|
|
![]() |
|
Figure 4: Desired laser position (cross-hairs) in the Setup view
Figure 5: Laser moved to cursor position.
|
Figure 6: Focus the optics on the cantilever.
![]() |
|
Return to Setup Procedures.